Compact pnCCD-based Colour X-Ray Camera systems for energy-dispersive X-ray detection in vacuum environments

CXC Core Vacuum Systems build on all features of the CXC Vacuum Systems platform and provide a compact, fully integrated mechanical design with detector and electronics housed in a single unit. They are designed for direct integration into vacuum chambers and synchrotron beamline environments where stability, simplicity, and minimal system complexity are key requirements.

These systems enable energy-dispersive X-ray detection under vacuum conditions, supporting reduced absorption losses, lower background, and improved access to low-energy X-rays. CXC Core Vacuum Systems are supplied as complete, ready-to-use detector instruments, including the pnCCD sensor, vacuum-compatible housing, cooling infrastructure, control electronics, and the PNCamSuite software environment.

Technical Specification

  Specification
Sensor Type Backside-illuminated pnCCD with frame-store architecture
Pixel Size 48 µm × 48 µm
Total Pixel Array 264 x 264 (Sensitive area)
264 x 528 (Sensitve area + framestore)
Sensor Geometry 12.3mm x 12.3mm active area
Sensor Sensitivity Range 0.1 to 30keV
Frame Rate 1,000fps (Full frame, continious readout)
Sensor Material and Thickness 450 µm fully depleted silicon
Operating Enviroment Ambient (in-air)
Connection Flexible umbilical between sensor head and core electronics
Control Software PNCamSuite for setup, acquisition, and analysis
Developer API (Python and C++)
Integration OEM instrumentation, laboratory and beamline use

Variants

  CXC Flex Ambient Pro CXC Flex Ambient Prime
Sensor Operation Type Pro Prime
Usage Profile For applications requiring optimal spectroscopic performance and maximum energy resolution. For flexible, easy-to-operate setups prioritising convenience and room-temperature operation, without the need for a Be window.
Best Energy Resolution ~145 eV (400 fps) @ Mn-Kα (5.895keV) ~180 eV (400 fps) @ Mn-Kα (5.895keV)
Energy Resolution
(400fps)
~145 eV @ Mn-Kα (5.895keV) ~180 eV @ Mn-Kα (5.895keV)
Reccomended Energy Range 0.2 to 30keV 0.2 to 30keV
Sensor Operating Temperature Cooled (approx. −20 °C) Room Temperature (approx. +20 °C)
X-Ray Entrance Window Vacuum-tight 50 µm Be
Windowless with optional protective film
Windowless with optional protective film
Vacuum Requirement Sensor head under vacuum Sensor head at atmospheric pressure (no vacuum required)

Looking for full specifications and mechanical drawings?

Request datasheet