CXC Flex Ambient Aperture Systems
pnCCD-based Colour X-Ray Camera systems with central aperture for energy-dispersive X-ray detection in ambient environments
CXC Flex Ambient Aperture Systems build on all features of the CXC Flex Ambient Systems platform and are designed for energy-dispersive X-ray detection in ambient environments where a central aperture is required. These systems support transmitted-beam, backscatter, and inline experimental geometries that require additional instrumentation downstream of the detector.
The pnCCD sensor incorporates a 2.5 mm central aperture, surrounded by a 4 mm insensitive region, enabling defined beam transmission while maintaining full spectroscopic performance across the remaining active area. A compact sensor head is connected via an umbilical cable to remotely located electronics, allowing flexible installation in confined, space-limited, or mechanically dynamic setups. Each system is supplied as a complete detector instrument, including cooling infrastructure, power supply, control electronics, and the PNCamSuite software environment for data acquisition and analysis.
Key Features
-
Builds on all features of the CXC Flex Ambient Systems platform, adding a central sensor aperture
-
Central aperture with a defined insensitive region, enabling transmitted-beam and downstream instrumentation geometries
-
Sensor head connected via an umbilical to remotely located core electronics for flexible installation
-
Suitable for backscatter, inline, and correlative measurement configurations
-
Full pnCCD-based, energy-dispersive X-ray detection outside the aperture region
-
Single-photon detection with per-pixel spectroscopic information
-
Full access to both analysed and raw detector data for transparent processing and re-analysis
-
Supplied as a complete system including detector head, cooling infrastructure, power supply, control electronics, and PNCamSuite software
-
Available in Pro and Prime variants to match different performance and operational requirements (see table below)
-
Suitable for a wide range of applications, including:
-
Backscatter X-ray fluorescence (XRF) measurements
-
Correlative XRF and transmission-based techniques
-
Inline X-ray detection with downstream optics or detectors
-
Custom laboratory and OEM X-ray spectroscopy setups
-
Further application examples are described on the Applications page
-
Technical Specification
| Specification | |
|---|---|
| Sensor Type | Backside-illuminated pnCCD with frame-store architecture |
| Pixel Size | 48 µm × 48 µm |
| Total Pixel Array | 264 x 264 (Sensitive area) 264 x 528 (Sensitve area + framestore) |
| Sensor Geometry | 12.3mm x 12.3mm active area with 2.5mm diameter central hole (4mm diameter insensitive region) |
| Sensor Sensitivity Range | 0.1 to 30keV |
| Frame Rate | 1,000fps (Full frame, continious readout) |
| Sensor Material and Thickness | 450 µm fully depleted silicon |
| Operating Enviroment | Ambient (in-air) |
| Connection | Flexible umbilical between sensor head and core electronics |
| Control Software | PNCamSuite for setup, acquisition, and analysis Developer API (Python and C++) |
| Integration | OEM instrumentation, laboratory and beamline use |
Variants
| CXC Flex Aperture Ambient Pro | CXC Flex Aperture Ambient Prime | |
|---|---|---|
| Sensor Operation Type | Pro | Prime |
| Usage Profile | For applications requiring optimal spectroscopic performance and maximum energy resolution | For flexible, easy-to-operate setups prioritising convenience |
| Best Energy Resolution | ~145 eV (400 fps) @ Mn-Kα (5.895keV) | ~180 eV (400 fps) @ Mn-Kα (5.895keV) |
| Energy Resolution (400fps) | ~145 eV @ Mn-Kα (5.895keV) | ~180 eV @ Mn-Kα (5.895keV) |
| Reccomended Energy Range | 1 to 30keV | 0.2 to 30keV |
| Sensor Operating Temperature | Cooled (approx. −20 °C) | Room Temperature (approx. +20 °C) |
| X-Ray Entrance Window | Vacuum-tight 50 µm Be | Windowless with optional protective film |
| Vacuum Requirement | Sensor head under vacuum | Sensor head at atmospheric pressure (no vacuum required) |
Looking for full specifications and mechanical drawings?