pnCCD-based Colour X-Ray Camera systems with remote electronics for energy-dispersive X-ray detection in vacuum environments

CXC Flex Vacuum Systems build on all features of the CXC Vacuum Systems platform and add a flexible mechanical architecture with remotely located electronics. A compact pnCCD sensor head is connected via an umbilical cable to the core electronics, allowing the detector to be positioned precisely inside a vacuum environment while keeping mass, heat generation, and service access outside the chamber where required.

These systems are designed for energy-dispersive X-ray detection under vacuum conditions in experiments where flexibility, space constraints, or complex geometries make a fully integrated detector impractical. Each system is supplied as a complete detector instrument, including vacuum-compatible detector head, cooling infrastructure, control electronics, and the PNCamSuite software environment.

Technical Specification

  Specification
Sensor Type Backside-illuminated pnCCD with frame-store architecture
Pixel Size 48 µm × 48 µm
Total Pixel Array 264 x 264 (Sensitive area)
264 x 528 (Sensitve area + framestore)
Sensor Geometry 12.3mm x 12.3mm active area
Sensor Sensitivity Range 0.1 to 30keV
Frame Rate 1,000fps (Full frame, continious readout)
Sensor Material and Thickness 450 µm fully depleted silicon
Operating Enviroment Ambient (in-air)
Connection Flexible umbilical between sensor head and core electronics
Control Software PNCamSuite for setup, acquisition, and analysis
Developer API (Python and C++)
Integration OEM instrumentation, laboratory and beamline use

Variants

  CXC Flex Ambient Pro CXC Flex Ambient Prime
Sensor Operation Type Pro Prime
Usage Profile For applications requiring optimal spectroscopic performance and maximum energy resolution. For flexible, easy-to-operate setups prioritising convenience and room-temperature operation, without the need for a Be window.
Best Energy Resolution ~145 eV (400 fps) @ Mn-Kα (5.895keV) ~180 eV (400 fps) @ Mn-Kα (5.895keV)
Energy Resolution
(400fps)
~145 eV @ Mn-Kα (5.895keV) ~180 eV @ Mn-Kα (5.895keV)
Reccomended Energy Range 1 to 30keV 0.2 to 30keV
Sensor Operating Temperature Cooled (approx. −20 °C) Room Temperature (approx. +20 °C)
X-Ray Entrance Window

Vacuum-tight 50 µm Be

Windowless with optional protecitve film

Windowless with optional protective film
Vacuum Requirement Sensor head under vacuum Sensor head at atmospheric pressure (no vacuum required)